5. Take an image of the
guide star with the CCD using the same procedure as in the alignment
section. First center the dome slit on the guide star and find it in
the Telrad

,
then center it on the camera with the computer (
Reflector setup must
use the finderscope

).
If you need to review the process, return to the
alignment
section.
6. When you have the guide star directly centered on the camera chip,
press
the left or right arrow to select either 1st star or 2nd star depending
on which side of the
meridian you
are on. West of meridian = 2nd star. East of meridian = 1rst star.
Note: The number
just refers to the side of the horizon. You will be aligning with only
one star.
7. Use the down and up arrows to select the name of the guide star that
you have in the
camera now. In the above example, the star is Alpha Cassiopeiae.
8. Press E to align. You will now be in the main screen.
5. Use the arrow keys to select an object by it's Messier number, NGC
number, or planet name. For example, if you were going to find NGC 7635
(The bubble nebula), you would first move the cursor (_) under 'Mes'
with the left and right arrows. Then you would press down so that the
NGC catalog is displayed. Finally, you would use the cursor (_) to
select 7635.
6. Once the object has been selected, press E for enter. The underscore
cursor should move to the left of the numbers. Once it is there, press
the up arrow to select relative position. This is indicated by an
asterisk.
7. Move the telescope so that the right ascension and declination are
zeroed out. Lock down the declination

first, then the right ascension

.
8.
(Refractor only)
Take a quick exposure of the object with the Focus Tools tab

.
(Reflector
only) Look through the
finderscope

.
If you are looking for a relatively bright object,
it should appear somewhere in the field of view as a gray blob. If you
are hunting for a very
dim object that cannot be seen right away, you can try a
longer
exposure or you can
adjust the background
and range. If the object is not visible, see the section on
star-hopping.
9. Center the object by using the hand controller

.
The section on aligning contains the
directional
information.
A more complete manual on the digital setting circles is available
here.
Autoguider Calibration
The camera has the ability to control the telescope so that it corrects
for any tracking errors that may occur in the mount. A separate CCD
chip, called the autoguider is used to follow a guide star.
At this point, you should have acquired an object to image. If it is a
bright planet or you are willing to take noisier short pictures, so you
may
skip this section.
1. Go to the Take Image tab in CCDsoft and select the autoguider radio
button

.
1x1 bin and auto-darked images are best.
2. Change the radio button back to Imager

.
Go to the Autoguider tab and change the settings to match the window
below. Click on Take Image.
3. Ideally, there should be many suitable calibration stars in the
frame. If not, try moving the telescope slightly with the hand paddle
or increasing exposure time.
4. Switch the hand controller cable and the relay cable so that the
computer controls the drive motors.
5. Click on the calibrate button to initiate. The X and Y direction
should move 20 seconds each.
6. If calibration is
successful, it will briefly say so in the status bar

.
If not, you will likely get a message that there was insufficient
movement

.
Try increasing the exposure time if you receive the error message.
Sometimes the
star can drift out of the screen during calibration. In this case, see
the
reflector
section on calibration star placement.
7. Click the Autoguide button. The telescope will now exactly follow
one star.
Calibrating
the
Autoguider - (Reflector
setup only - click
to skip)
1. Go to the Take Image tab in CCDsoft and select the autoguider radio
button

.
Match your settings with the window below. 3x3 bin and auto-darked
images are best.
2. Change the radio button back to Imager

.
3. Look through the finderscope. If the eyepiece illuminator is
on

,
you should see some red crosshairs.
The following diagram shows approximately where the autoguider is
inside the finderscope eyepiece

.
4. Find a suitable calibration star in the eyepiece

.
It should be
completely BY ITSELF with no other stars as bright nearby. Almost any
star that can be seen in the finderscope will do, but the brighter the
better. Using the hand controller

,
move the telescope so that a suitable guide star is inside the
autoguider area

.
5. In CCDsoft, click on the
Autoguide tab in the Camera Control window. Make the settings match the
window below. Click Take Image. A bright star should be visible in the
autoguider frame. If not, keep trying to place the star where it should
be in the finderscope eyepiece

or increase exposure time if needed.
Another method for placing the
guide star is available in the
autoguide
trouble appendix.
6. Once the star is in the autoguider frame, switch the hand controller
cable and the relay cable so that the computer controls the drive
motors.
7. Depending on if you are
on the
western or eastern side of the
meridian, put the
star near the
appropriate corner, but not too close, because it will move during
calibration.
8. Click on the Calibrate button. Set the X move and Y move to 6
seconds. The star will move during the process.
If calibration is successful, it will briefly say so in the status
bar

.
If not, you will likely get a message that there was insufficient
movement

.
This is the standard error if anything goes wrong. Try again with
longer exposures, different stars or see the appendix on
autoguider
trouble.
9. If the calibration is successful, the software will briefly say so
in the status bar of the Camera Control window

.
Move the telescope back
to the object you wish to image. To do this, you may have to plug the
hand controller back in

.
10. Look in the eyepiece of the finderscope

.
If there is a star right in the autoguider area, you are in luck -
skip to step 13.
Otherwise, search the nearby area for a suitable guide star. If none is
available, you can try the alternative method in the
autoguider
trouble appendix.
11. Switch the hand paddle and camera relay wires again so that you can
use the hand paddle.
12. Looking through the
finderscope, move the autoguider
area

to a suitable guide star with the hand paddle

.
13. IMPORTANT -
Remember to plug the relay cable in

if it has been unplugged.
Click on the Take
Image button to make sure the guide star is on the chip. If it is,
click on the Autoguide button to begin autoguiding.
Grayscale Imaging
By now, you should have acquired an object to image and
started the autoguiding. The last step is to take the images
themselves. The procedure from this point is flexible and depends
highly on what you are looking at. Most likely, capturing great
exposures will require
a little playing around. The following guidelines will help, but feel
free to adapt the settings to your needs.
1. Tag your images. The Setup tab has an option for File Defaults. Type
in your name and the telescope information. The following shows the
information for the reflector. The refractor should have the following
settings: Description="Refractor", Focal Length="600", Aperture
area="7854", Aperture diameter="100"

2. Auto-save your images.
Switch to the
AutoSave tab. Everything should match the window below except the File
name prefix name and the folder to save it in. The convention for
saving
files is to put them in a folder with the date. For example, images
taken on 8/18/06 should be saved in "C:\Documents and Settings\ccd\My
Documents\CCD images\8.18.06"
3. Capture your images
This is the main image-capture window, where most changes will be made:
EXPOSURE (Minutes,
Seconds): Here is
where you will set your exposure length. The main idea is to take
as long an exposure as possible. The more signal you have, the greater
the signal to noise ratio and the better the picture; however, if
exposures are too long, you will begin to see blooming

.
Try to take exposures that come close to blooming, but don't exceed the
maximum pixel value (saturation). If you hold the cursor over a
star, the pixel value is displayed in the lower left corner.
They
should range in value from 1 to 65,535. Sometimes, blooming will be
necessary in order to image faint nebulosity.
For the refractor, 1 minute exposures
are generally
long enough to capture good detail. Taking several images and stacking
them up later will allow total exposures of 10 minutes or more.
For the C14 reflector, longer exposures
are usually
necessary. Galaxies and nebulosity usually need more than five minutes
to achieve good signal to noise ratios.
Planets should be imaged with as short
an exposure as possible (0.1 seconds is as fast as our camera goes).
DELAY (seconds): Sets the time the camera will wait before opening the
shutter. Often, it is useful to set a 15 second delay before an image
to give enough time to go downstairs and stay away from disturbing the
system during an exposure. If you choose to sit and watch the images
right at the computer, set the delay to zero.
SERIES OF: This option allows several images to be taken in a row. When
used, the Take Image button changes to a Take Series button.
FILTER: Generally, the luminance filter should be used to take images.
Color shots will be described in the next section.
TO NEW WINDOW: Leave off, unless you want to compare images.
SUBFRAME: Leave off unless you are focusing.
BIN: For high-resolution images in the
refractor, 1x1
binning is best. However, with the
reflector,
3x3 binning is better for two reasons: the seeing conditions make high
resolution images unnecessary and 3x3 binning will be brighter.
IMAGE: Set to Light frame and set Reduction to None. The image
reduction will occur later in post-processing.
Imager/Autoguider radio button: Use the Imager.
Take Image/Take Series button: Click here to begin the exposure.
If the image is too large
for the screen

,
press Alt-3 to make it smaller

.
(Control 2,3, and 4 zoom in while Alt 2,3,4 zoom
out. Control-1 = Alt-1 = normal size.)
Important note about image capture:
The telescope is
highly sensitive to any small vibrations. If you simply walk around the
pier, you can ruin an entire 10 minute image.
This
means you should tiptoe
and move very slowly during an exposure. Another option is to set the
delay to 15
seconds and
leave the area during the delay before exposure time. Feel free to gaze
at the stars
from the deck in the interim.
One thing to occasionally check up on is
dome cutoff. Keep the dome slit aimed

so that the Telrad target

stays approximately in the center of the dome slit. Unlike
walking on the floorboards, this vibration does not ruin images.
Color Imaging
Color pictures are
made by assembling images taken through luminance, red, green and blue
(LRGB) filters. You need to capture separate exposures through each
filter and combine them later. The camera will be most sensitive
through the luminance (no blockage) and least sensitive through the
blue filter, so exposure length may vary.
1. Read the
grayscale
imaging section first so you set your tag and autosave
information.
2. Switch to the Color Tab to set the color settings.
3. Change each menu by
clicking on the
Edit button. You will be prompted with a small window in which you can
change the settings. Leave the Filter options as they are. For the
refractor, 1
minute exposures through each filter work well. For the
reflector,
10 minute exposures through each filter work well. However, you may
adjust the times to favor the blue exposures. If the times are all
different, it can complicate taking dark frames (
next section). For
most situations, take 1 through each filter. Set the bin to
1x1
for the refractor
and to 3x3
for the
reflector.
4. Click on the Take Color/Take Series button to begin. The camera has
a built-in color wheel that will change filters between shots.
Capturing Dark Frames
The images you have
captured so far probably contain a salt-like smattering of bright
pixels on top of the galaxy/nebula/star cluster. Most of these pixels
are the result of
dark
current, not distant stars. You can subtract out these pixels
by taking dark
frames.
1. Switch back to the Take Image tab and select Dark frame.
2. If you took grayscale images, click on Take Image to capture dark
frames with equivalent exposure time and Bin mode. If you took color
images, click on Take Color in the Color Tab. IMPORTANT: Make sure you
have exactly the same exposure
time and bin mode as before. Old images can be re-opened and 'View
| File Information' will tell you the duration of the exposures. As
they
are taken, the Status bar will show the progress.
3. Take several dark
exposures, so they
can be averaged. If they are 10 minutes or longer, you can get by with
only one, but shorter exposures, like 1 minute should be taken thrice
and averaged.
Cleanup
If you are still interested in taking more shots and it isn't too late
in the morning, return to
selecting
and object.
It will be necessary to re-calibrate the autoguider, but dark frames
may be re-used if they are the same exposure length. If you are
exhausted and ready for bed, proceed on.
You may turn on the white dome lights for illumination

.
1. Close all software programs. They will automatically disconnect the
devices.
2. Turn off the camera power with the I/O switch

.
3. Turn off the Telrad

by fully closing the dimmer switch.
4.
Reflector only (Click
to skip). Turn off the Robofocuser

.
5.
Reflector only.
Turn off the reticle illuminator

and store it in the eyepiece
cabinet

.
6.
Refractor only -
Disconnect the Robofocuser and coil up the wires.
7.
Disconnect the camera connections one by one. Coil up the
power chord

,
relay box

and USB cable

.
Connect the
hand paddle

if it has been unplugged.
8. "Park" the telescope by pointing it upright and north

.
9. Remove the dew shield

and put the lid back on

.
10. Lock down the right ascension

and declination

knobs.
11. Untie the camera's strap

.
12. Loosen the set screws (Refractor

Reflector

)
that hold the
camera and carefully remove it from the telescope. Put the camera back
in the padded case.
13. Replace the eyepiece where camera was (Refractor

Reflector

).
14. Loosen the right ascension

and declination

knobs.
15. Retrieve the dust cover and put it back on the telescope with the
strings hanging on either side of the counterweights

.
16. Return the dome so that it faces directly east

.
17. Close the dome shutters with the filter wheel

.
18. Close and lock each window and door behind you. Be sure to turn off
all lights.
Image Reduction
Post processing is best done in the
daylight, but if
you are eager to see some color shots, you may proceed anyway. The
following procedure uses CCDsoft, but MIRA or any image-processing
program would work as well.
1. Organize your images. The following diagram gives a good folder
structure to follow. It is important to have the images labeled by
their name and to know where the flat frames and dark frames are. If
there are many different exposure times, it is useful to label the dark
frames with the exposure duration in the title.
2. Click on "Image | Reduce | Image Reduction ..." to open up the
reduction window.
3. Click "+" on a reduction group (any will do) to open up the frame
folders.
4. Delete any old images in that reduction group.
5. Click on Dark Frames and then click the Add Frames button. In the
selection window, select all relevant dark frames (those with
equivalent exposure time to the image you are working on).
Control-clicking will allow you to select multiple frames. Do the same
for Flat-frames, keeping in mind a specific filter. Skip bias frames
and Darks for Flats only.
Why?
Note: Multiple frames are averaged to increase the signal to noise
ratio.
Note: If you do not have flat-frames, delete any old ones with the
Remove button.
5. Once the relevant frames are loaded, use the Reduce or Reduce Folder
command to reduce an entire folder of images. If a whole folder is to
be
reduced, nothing else can be inside that folder.
6. Repeat steps 3-5 for all images, treating each filter separately.
Adjusting Background and
Range
Raw CCD images have millions of pixels.
Each pixel can range in value from 1 to 65,535. Often, most of the
detail is contained in a range of a few thousand brightness values.
Honing this area is easy and non-destructive. We simply adjust the
black point and white point, which set the maximum and minimum
brightness. This technique, along with being easy, is one of the most
effective tools for enhancing faint nebulosity.
1. Open up the desired image.
2. Open the histogram editor by clicking "Image | Brightness &
Contrast | Histogram..."
3. Adjust the green sliders until the desired detail is reached.
Putting the white point (left triangle) at the first peak of dark
pixels

tends to help cut out any background noise.
4. Save the image when you are done. You can always adjust the
histogram again without destroying anything.
Stacking Multiple Images
If you wish to combine many short
exposures, you can simply add pixel values. Do not use this for
photometry, unless the pixels stay below 65,535 counts.
Note: this should only be done after images have been reduced.
1. Place all images to be stacked in a folder with no other files.
2. Select "Image | Align | Align Folder of Images"
3. Select the correct folder.
4. Select the folder to place the aligned images in. You should create
another folder in a similar spot, but it needs to be outside of the
un-aligned shots.
5. Click Align.
6. Select "Image | Combine | Combine Folder of Images"
7. Select the folder that contains the aligned images.
8. Select the Add function.
9. Click Combine.
Nonlinear Stretches
Sometimes, you may find that adjusting
the background and range cuts out faint detail or saturates the bright
cores of galaxies. For these situations, it is useful to try the
Mid-Range Brighten. This tool "evens out" the image so that you can
display dim outer structures without over-doing the brighter regions.
1. Click on "Image | Brightness & Contrast | Mid-Range
Brighten". The modified image will open in a new
window, leaving the original untouched.
Color Combining
For color shots, you need to have at
least one
exposure through red, green, and blue filters. A luminous shot is
desirable, but not necessary. If you have multiple shots through each
filter, they should be combined, as described in the
multiple image combination
section. This leaves you with a master luminance, red, green
and blue.
1. Align the images as described in the the
multiple
image combination
section. Open them with CCDsoft.
2. Select "Image | Color | Color Combine..."
3. In the new window, make sure that the red, green, blue and luminance
(if you have one) selections match the appropriate shots.
4. If there is no color preview, click on the Preview button. It may be
necessary to shrink this window by pressing Alt-3.
5. Adjust the color sliders until the majority of the stars are white.
You can also compare your image with published ones to reach the
appropriate color balance.
6. Click on Combine.
7. Repeat steps 5-6 several times until you have a few versions.
Compare them and choose the best.
Note: You may wish to
adjust the background
and range
of the individual filter shots first. Modifications to the luminance
affect results more than modifications of the red, green or blue.
Photometry
The magnitude of stars can be measured
by using the
CCDsoft program, or any similar CCD editor. You will need to know the
magnitude of at least one star in the image in order to proceed. This
can be found in any star catalog, but it requires that you know where
in the sky the image was taken. TheSky, a planetarium program, has the
capability to align your image with its catalog.
1. Find the magnitude of a particular star in the image. It should not
be a variable.
2. Click on the set magnitude button.
3. Click on the star that
will be calibrated against. Type in the appropriate magnitude.
4. Measure the magnitude of stars, quasars, asteroids or comets with
the determine magnitude button.
Balance
One common balancing problem is electrical chords. Check first to make
sure this is not the issue. If balance is off, you must figure out if
it is in the right ascension axis

,
declination axis

,
or both.
1. If the declination

and
right ascension knobs

have been tightened, turn them counter-clockwise
until they feel loose.
2. Move the telescope with the two D rings

located on the sides.
Point it directly West. If the telescope moves when you
let go, it is imbalanced in declination.
3. Rotate the right ascension axis so the counterweights bar is
horizontal. If the moves from the horizontal position, it is imbalanced
in right ascension.
4. IMPORTANT - Lock down the right ascension

and declination

knobs. You don't want things swinging around as you adjust them.
5. If there is a small amount of imbalance on the right ascension
axis

, you may
compensate for it by moving back and forth the small brass weights on
the end of the telescope

.
Do NOT move the large counterweights unless it is absolutely necessary.
If it is necessary, turn on the dome lights. To move a large
counterweight, loosen the lock knob slightly and screw the
counterweight in the desired direction. Do not move the counterweight
with the brass ones attached.
Check to make sure all lock knobs for counterweights are secure before
proceeding.
6. If there is imbalance in the declination axis, you have to decide
whether it is tolerable or not because this procedure can be very
tricky: Please read through it before beginning.
a. Remove the camera so that if the
telescope drops, at least the camera is not destroyed. (See
cleanup for procedure)
b. Brace your arm and shoulder against
the telescope back with all your might (you will be holding up about
60% of the weight of everything but you will have friction to help you).
c. Loosen the two lock knobs so that you
can slide the telescope forward or backward to change the weighting
7. Retest the balance and make further adjustments if necessary.
Fixing Poor Alignment
Misalignment can take 20 minutes or more
to fix, but should not have to be performed often.
1. Find a really bright object in the sky, preferably a planet like
Jupiter, but a bright star will work as well. Avoid using the moon.
2. Prepare the camera to take continuous, short exposures
as described in the alignment
section.
3. Make sure the right ascension

and declination

knobs are loose. Point the telescope's Telrad

at the bright object.
4. If the camera does not record the planet or star, make circles
around it by pointing the telrad. Start with small circles, and move
outward while watching the computer the whole time. When the object is
in view, it will probably streak across the screen.
5. Adjust
the telrad as in the alignment section,
step
9 so that the target centers the star

.
6.
(Reflector setup only)
If the finderscope is off, re-center the star/planet as in the
alignment section,
step
10.
Finding Focus - Reflector
Setup
If the focus is way off, so that stars
are
unreasonably large or have holes in the center, it is probably
necessary to take out the camera and make some adjustments.
1. Remove the camera as described in the cleanup section,
step 11. You can
leave the power on and set it in the padded case, as long as you are
careful and put on the dust cap

.
2. Attach the ground glass

in place of where the camera goes.
3. Point the telescope at a bright star (Not a planet!), making sure
the right ascension

and declination

knobs are loose.
4. There should be a disk visible on the ground glass, unless it is too
far out of focus. Turn the SCT focus knob until the star becomes a
point. You may have to turn it far to observe a change in size of the
disk.
5. When the star has approached a point on the ground glass, it should
be very close if not exactly where the camera should focus. Proceed
with the
normal
focusing procedure.
Autoguider Trouble
Setting up the autoguider can be one of
the most
frustrating parts of CCD work at the Peters Observatory. If you are
short on time or patience, you might want to try short exposures (15
seconds in the
reflector
and 45 seconds in the
refractor)
without autoguiding.
Calibration Problems
The standard message for any kind of
problem is that
there was insufficient movement in the X or Y direction. This error
sometimes describes the problem and sometimes not. Here are some tips:
- Choose another guide star (ie. move the telescope).
Sometimes the software gets confused as to which star it should follow.
- Increase the calibration time to be sure that the star is
registering.
- If you click on the Auto button, the computer will flash a
box quickly
in the region that the star calibrates on. If this box does not
enclose a star, you need a longer exposure or a brighter star.
- Make sure the relay wire is plugged in
.
Without this
cable, the computer has no control over the telescope.
- Place the calibration star in the appropriate corner,
so that it does not move off of the screen.
- Make sure that the Losmandy Drive System settings are optimal
- Occasionally movement is actually too small. Increase the
calibration time slightly.
- Depending on the refractor or reflector setup, make sure
the
autoguider binning mode is correct (refractor --> 1x1, reflector
--> 3x3)
Alternative Star Placement
There is another method for
placing a guide star, whether for calibration or guiding.
1. Find a nice bright star in the imager window (the main CCD
window). It will require the least work if this star is in the top
middle.
2. Use the hand paddle (see directionality in the
alignment section)
to move the star straight off the top-middle of the imager window. Go a
few seconds beyond until it shows up in the autoguider window.
Autoguide failure
If the autoguiding
fails midway through an exposure, it is usually because the camera has
paused and the star has drifted in the interim. Here are some tips:
- If you are taking color exposures, the blue filter can be
so dim
that the star is difficult to find. Try increasing autoguider exposure
time.
- If the star is too dim, you may need to find another or
increase
exposure time. Hovering the cursor over the star will give you an idea
of the pixel count. Above a value of 1000 is best.
- To see which star the software was following, click on the
Auto
button. This should draw a box around the star. If it does not follow
the star you intended, simply click on the correct star. You should see
a box flash around where you clicked and the computer will then follow
that star.
- Sometimes, when the equatorial mount is out of alignment,
the
star drifts too fast to follow. Fine-tune polar alignment as described
in the appendix.
- If need be, as a complete last resort, you can try turning
the camera so that a favorable star lands on the autoguider.
Unfortunately, this means re-focusing and re-calibrating (at an angle).
To do this, loosen the camera's set screws and face the camera wires in
the direction of the guide star. Note that directionality of the hand
paddle will be rotated.
Refining Polar Alignment
If the telescope does
not follow the stars well, it may be that polar alignment is off. You
can use the camera to help fine-tune polar alignment. Read the whole
procedure before executing becauset there are a few quick moves that
you need to make. The following is based off a procedure from
http://www.ucihs.uci.edu/pandb/hall/polar.htm.
Altitude Refinement
1. Point the telescope to
either the East or West and find some relativley dim stars,
magnitude
6 or fainter. You will be creating star streaks, so smaller stars will
make thinner lines.
2. Start a 65
second exposure, as described in the
take
image section.
3. After 5 seconds, turn off the losmandy drive system, remembering to
ground your hand.
4. Wait 30 seconds, then
turn on the
drive system again. Making sure to return the settings to 2X. Return
the star near its original position by holding the hand paddle

button
for another 30 seconds.
5. The image will tell you
how far off
the alignment is. If the altitude alignment is perfect, you will see
one line. If it is not, you will see a V. The following diagrams were
taken from
http://www.ucihs.uci.edu/pandb/hall/polar.htm.
6. If there is a V, your goal it so make it a single line. Start by
adjusting the altitude knob slightly in one direction. You may have to
loosen the lock knob

first.
7. Repeat
steps 2-6 again,
noting if the V becamer narrower or wider. Keep adjusting until you
approach a straight line.
Azimuth Refinement
1. Point the telescope at a
6th
magnitude
or dimmer star near the zenith (straight overhead).
2. Repeat
steps
2-5, but this time you will adjust the azimuth knobs

.
You may need to loosen the left

and right

set screws first.
3. Keep adjusting until you
get a straight line.
Using the Planetarium
Software
There are many planetarium programs loaded on both the Linux and
Windows computer. They include Hallo Northern Sky, Starry Night Pro,
TheSky and KStars. The following procedure shows how to use Starry
Night Pro as a celestial atlas.
1. Open Starry Night Pro
from the desktop.
2. Search for the object you are interested in finding by pressing
"Control-f" and typing the name in. Hit Enter.
The object should appear in Starry Night
Pro with a label.
3. Click OK in the registration window.
4. Go to the FOV tab and select the 2º and 4º
checkboxes under Other (All Charts).
Refractor: Click on SBIG ST10XME under Refractor

.
Reflector: Click on SBIG ST10XME under C14

.
5. You will now see a box and two circles. The circles represent
the Telrad target

(4 degrees and 2 degrees). The box represents what the camera sees. Use
this box to gauge how big the object will appear in the final image.
Refractor
Reflector
The Basics of Star-Hopping
Before the days of settings circles and automatically-pointing
telescopes, amateur and professional astronomers used to resort to sky
charts to find celestial objects. The following procedure gives a rough
idea of the process using the Starry Night program as a guide. It is
simple in principle, but can be very frustating and tiring in practice.
1. As described in the
previous
section, set up the software to show you how big the Telrad
circles will be.
2. Find a bright constellation near the object you are looking for. You
may want to click on the stick figure option

to see the constellation figure.
Find it both in the sky and on the computer.
3. Using the arrow keys, center the object in the two
Circles.
4. Using the position of those circles as a guide, align
the Telrad target

in the same way. This will give close alignment.
5.
Reflector Setup Only
(Refractor setup can use the camera if necessary).
For futhur alignment, zoom in closer with the computer. Flip the x and
y axis so that it will appear as seen through the scope.
Look through the finderscope eyepiece

and identify some stars that are seen on the computer. For example, you
may notice the curved line of stars highlighted below. Make any
additional movements to find your object of interest.
CCD Acronyms
ADU - Analog to digital unit
CCD - Charge Coupled Device
DSC - Digital setting circles
FITS - Flexible Image Transport System (image format)
FOV - Field of view (optical or for a pixel)
FWHM - Full width half maximum (width of 50% of brightest starlight)
JPEG - Joint Photographics Expert Group (image format)
LRGB - Luminance layer, red, green blue
NGC - New general catalog
PIXEL - Picture Element
PSF - Point spread function
SCT - Schmidt Cassegrain Telescope
SNR - Signal to Noise ratio
TE - Thermo electric (cooling)
TIFF - Tag image file format
UGC - Uppsala General Catalog
Written by Everett
Schlawin